Data sheet

HUMOR20

The role of humidity calibrations that are accurate, reproducible, and documentable is becoming more and more important. ISO quality guidelines and regulations according to FDA guidelines in the pharmaceutical industry, etc., require that humidity instruments have a traceable, accurate calibration. The humidity calibrator HUMOR 20 developed by E+E is the ideal reference instrument for these requirements.

The HUMOR 20 can be used in the humidity range of 10-95% RH both for monitoring cylindrical sensors (transmitters, hand-held instruments,...) and also for monitoring instruments with cubic dimensions (data loggers, wall instruments,...). A temperature sensor integrated in the measurement chamber also permits the monitoring of an optional temperature output.

The HUMOR 20 is traceable to international standards and can be delivered with an official, internationally recognised OEKD calibration certificate. Due to its high accuracy, the HUMOR 20 is the basis for accredited calibration laboratories for relative humidity.

Based on its operating principle, the HUMOR 20 can be used under typical conditions in a laboratory climate. This means that expensive, fully air-conditioned rooms are not necessary. For operation HUMOR 20 requires only distilled water, filtered oil-free air with a pressure of 10 bar and a power supply between 90-230 V AC. The specimen can be powered by 24 V DC that is available directly on the HUMOR 20.

  • Highest accuracy

  • Traceable calibration

  • Independent of ambient temperature

  • Easy handling

  • Traceable to international standards

  • OEKD certificatable

  • Function principle ___________ two-pressure-reactor

  • Working range _____________ 10...95 % RH

  • Protection class ____________  I

  • Protection type _____________ IP40

  • Surge voltage category  ______ II

  • Installation altitude __________ up to 2000 m above sea level

  • Application ________________ Indoors