Display reads oxygen from 0.01% up to 25.0% with no range changes
Analog output and alarms can be configured to operate over any of ten ranges from 0-1% to 0-25%
Analog output 4-20mA, isolated.
Complete sample system built into AMI’s patented cell block.
Easily replaceable sensor, no tools required.
Very rapid response time from air to low percentage levels.
P-2 sensor standard, other sensors optional.
Simple, versatile installation.
Operates off 117VAC with very low power consumption
Analog output can be easily calibrated to an external device.
2 independent, fully adjustable alarm settings with relay contacts.
Complete alarm logic programming: latching or non-latching, open or close on alarm, high alarm or low alarm, alarm-on delays and alarm hold-off.
Integral data logger: Logs data for 15 days @ 1min intervals and 30 days @ 2min. intervals, etc.
USB connectivity to a PC: Allows complete access to the internal functions and settings.
ModBus Industry standard protocol over RS48
Oxygen Sensor life indication.
4 user selectable output ranges
3 ½ digit LCD
2 fully adjustable oxygen concentration alarms
Alarm hold off/Bypass
RFI protected
4-20mA. isolated analog output
Analog output calibration synchronizes the analog output with an external monitoring system
Data logger
Fully programmable alarm delays
Sensor calibration history
USB port for configuration and access to advanced features
Modbus using Bidirectional RS485 for industry-standard communications
Low minimum detection limit 0.01%
Excellent repeatability
Extended operating temperature range
Fast upscale/downscale response times
Patented Cellblock Technology: Allows for all components such as flow control valve, flow meter, Sample, span and seal-off solenoid valves and compression fittings to be an integral part of the cellblocks, eliminating long lengths of tubing and fittings.
Benefits of this design include compact size, faster response times and front panel sensor access without the need for tools.
Unaffected by changes in flow rate from 0.1 to 2.0 SCFH
Panel mount
Compact size
2 year warranty for analyzer, parts and labor
6 month sensor warranty, life expectancy 1-2 years
These items require the AMI User Interface Program
201R Series Standard ranges:
0 – 1%, 0 – 5%, 0 – 10%, 0 –25%
Optional ranges: 0-1000ppm, 0-5000ppm, 0-50%
Sensitivity: 0.5% of full scale
Repeatability: +/- 1% of full scale at constant temperature
Operating temperature: 41°F to 113°F.
Maximum inlet pressure: 40psig
Humidity: < 85%, non-condensing
Operational conditions: Pollution degree 2, Installation category I I.
Drift: +/- 1% of full scale in 4 weeks at constant temperature
Expected cell life: 9 months to 2 years.
Response time:
90% of full scale in less than 10 sec (depending on sensor)
Output: 4-20mA isolated.
Alarm contacts: 230/117VAC @ 5A, or 28VDC @ 5A, resistive
Datalogger Logs data for 15 days @ 1 minute intervals; 30 days @ 2 minute intervals, etc. Represents the output range selected: 0-10ppm, 0-50ppm, 0-100ppm, 0-500ppm, 0-1000ppm, 0-.500%, 0-1%, 0-5%, 0-10% and 25%. Overange capability up to ten times the range selected.
Power requirements: 90-240VAC <10W.
Absolute Maximum Power voltage 264VAC 60Hz.
¼” stainless steel compression fittings for sample, span gas and exhaust.
Overall dimensions: 9” w x 5” h x 7” d
Weight 10 lbs